Manufacturer and Exporter of Blood Bank Refrigerator

Blood Bank Refrigerator

The Maxcare Blood Bank Refrigerators are designed to ensure consistency in performance and uniform temperature for all the units of Whole Blood and Packed Red Cells in storage by utilizing Maxcare efficient technology.

Maxcare Blood Bank Refrigerators with a highly efficient compressor that provides superior temperature recovery, rapid cooling and quiet performance for each model. Designed with all the features for blood product preservation such as superior uniformity, microprocessor control, and large flexible storage capacity. Whatever your preservation needs are, Maxcare provides the right equipment to meet your exact needs.

The blood bank refrigerator are critical components of blood banks, hospitals, and other healthcare facilities where blood are stored and used. A blood bank refrigerator is a refrigerator manufactured to store blood products at optimal temperatures to maintain their viability and safety for transfusion.

Silent Features
Blood Bank Refrigerator
  • 7-inch-wide Touch screen operating LED/LCD display (Which can be used with Gloved hand).
  • Fully programmable microprocessor controlled/PLC base- controller with membrane keypad and eye level control panel.
  • In built facility for User login, User ID tracking, run logs, protocol tracking with RFID/Password.
  • User access control and security with RFID/Password Protected security lock on the control panel for unauthorized tampering.
  • On-board real-time data logging, connectivity via RS 485 port data management software.
  • Displays both set and actual run conditions simultaneously.
  • Internal memory for storing temperature & Alarm data records which can displayed and transferred to PC.
  • Digital Temperature display and Recorder with provision for SMS alert and remote monitoring and controlling system.
  • °C C/F Temperature Choice.
Accessories
Key features of Blood Bank Refrigerators include:
Silent Features

Maxcare Make Blood Bank Refrigerators

Model MaxBBR-60 MaxBBR-120 MaxBBR-200 MBBR-300
Type Vertical Vertical Vertical Vertical
Total Storage Capacity 060 Bags Litres: 155 120 Bags Litres: 228 200 Bags Litres: 442 300 Bags Litres:655
Blood Bag Capacity per Tray 450ml-60 Nos. 350ml-80 Nos. 450ml-120 Nos. 350ml-160 Nos. 450ml-200 Nos. 350ml-260 Nos. 450ml-300 Nos. 350ml-510 Nos.
Trays 04 Nos. Slide out SS 304 trays with removable fixed gap Acrylic Panel. 04 Nos. Slide out SS 304 trays with removable fixed gap Acrylic Panel. 05 Nos. Slide out SS 304 trays with removable fixed gap Acrylic Panel. 05 Nos. Slide out SS 304 trays with removable fixed gap Acrylic Panel.
Power Consumption 550W 550W 550W 700W
External Dimensions (WxDxH) mm. 600x688x1191 600x688x1481 760x788x1801 860x948x1887
Internal Dimensions (WxDxH) mm. 500x500x620 500x500x910 600x600x1230 700x760x1230
Weight 137Kg 165Kg 205Kg 265Kg

Technical Specifications of Blood Bank Refrigerator

Make Maxsol Instruments
Outer Material of Cabinet Corrosion resistance CRCA/GI powder coated steel (18swg)
Inner Material of Cabinet Stainless Steel 304
PUFF Insulation-CFC free urethene
  • Cabinet: 80 mm
  • Door: 80 mm
Ground Clearance 4 castor wheels (2 Lockable and 2 un-lockable) for free mobility.
Door Single/Double (optional) Integrated Door handle and latch assembly to allow for easy operation, and includes standard key lock and provision for padlock.
Inner Door Separate acrylic sheet inner door which prevents cold air loss during door opening.
Compressor Hermetically sealed industrial grade,high efficiency air cooled compressor.
Refrigerants CFC Free-HCFC Free non flammable green refrigerants.
Circulation inside the Chamber Continuously operated forced Uniform air circulation using the evaporator fan
Pull down time 04 Hours
Temperature Controller Fully programmable microprocessor controlled controller with membrane keypad and eye level control panel.
Display with Resolution 4x7 Segment LED with 0.1°C resolution. °C
Operating Temperature Range +2°C to +8°C adjustable with setting accuracy +0.1°C
User Parameter Setting Temp. Set point
  • High alarm
  • Low alarm
  • Buzzer off
Temperature uniformity ±1˚C or Less
Environmental Condition -10°C to +45°C with Humidity 5% to 95% RH
Temperature Hold Over In case of power failure the rate of fall of temperature not more than 2°C/Hour with the holdover time of 2 Hours.
Temperature Monitoring System With IOT (Cloud Manage) supported (Optional) and 7 days inkless Chart Recorder
Audio Alarms and Visual Indications
  • Temperature High
  • Temperature Low
  • Door open
  • Power failure
  • Sensor failure
Visual Indications
  • Line in
  • Power On
  • Heater ON
  • System OK
  • Battery Low
  • Battery ON
  • Chart change
  • Compressor On
Battery Backup 12 Hours Battery Backup for Digital Display, Total Alarm system, Temp. recording system for TRCU ensures uninterrupted temperature tracking with auto recharging facility.
Noise level <= 55 db
Power supply 220VAC, 50 Hz +10%
Certificate